氣(qi)路(lu)檢(jian)查(zha),氣路(lu)是(shi)安全(quan)操(cao)作(zuo)的關(guan)鍵(jian)。
環保(bao)測(ce)試(shi)光譜(pu)儀(yi)所用的燃(ran)氣(qi)為乙(yi)炔,它與空(kong)氣(qi)混合點(dian)燃(ran)易發生(sheng)爆炸(zha)。如(ru)果(guo)出(chu)現泄漏(lou),將威(wei)脅(xie)操(cao)作(zuo)者(zhe)的人(ren)身(shen)安全(quan)。因此(ci)氣密(mi)性的檢(jian)查(zha)是(shi)要任(ren)務(wu)。每次(ci)使(shi)用之(zhi)前要(yao)檢(jian)查(zha)電路及氣路的連(lian)接。
絕多(duo)數(shu)儀(yi)器的氣(qi)路(lu)通(tong)常(chang)采用聚(ju)乙(yi)烯(xi)塑(su)料管(guan),時(shi)間(jian)長(chang)了(le)非(fei)常(chang)容(rong)易老化(hua)(建(jian)議聚(ju)乙(yi)烯(xi)塑(su)料管(guan)使(shi)用年限2~3年)。因此(ci)要經(jing)常(chang)對(dui)氣(qi)路(lu)進行(xing)檢(jian)測、檢(jian)漏(lou),尤其(qi)是(shi)乙(yi)炔氣(qi)滲(shen)漏(lou)可(ke)能(neng)造(zao)成(cheng)事(shi)故(gu)。嚴格(ge)禁(jin)止在乙(yi)炔氣(qi)路(lu)管道(dao)中(zhong)使(shi)用紫銅(tong)、黃(huang)銅及銀制(zhi)零(ling)配(pei)件,並要嚴禁(jin)使(shi)用油(you)類(lei)物質,測試(shi)高(gao)濃度銅或(huo)銀溶液(ye)時(shi),應(ying)反復用去離(li)子(zi)水(shui)中(zhong)噴洗(xi)。用(yong)棉簽(qian)沾(zhan)肥(fei)皂水(shui)塗(tu)抹(mo)氣路各連(lian)接處(chu),遇(yu)有漏(lou)氣現象要立即(ji)更換(huan)元(yuan)件。檢(jian)測完(wan)畢(bi)後壹(yi)定要(yao)將(jiang)肥(fei)皂水(shui)擦(ca)拭幹(gan)凈。當儀器測定完畢(bi)後,要(yao)先(xian)關(guan)閉(bi)乙(yi)炔鋼瓶(ping)輸(shu)出(chu)閥(fa)門(men),其(qi)次(ci)等(deng)燃(ran)燒(shao)器上火焰(yan)熄(xi)滅(mie)後再關避(bi)儀(yi)器上的燃(ran)氣(qi)閥(fa),後再關避(bi)空(kong)氣(qi)壓(ya)縮機(ji)的開(kai)關(guan),以確(que)保操(cao)作(zuo)者(zhe)的人(ren)身(shen)安全(quan)。
檢(jian)查(zha)廢(fei)液(ye)管中(zhong)部(bu)要有水(shui)封,避免回(hui)火現象發生。同時(shi)廢(fei)液(ye)管出(chu)口(kou)處(chu)不(bu)要插入(ru)廢(fei)液(ye)桶(tong)液(ye)面(mian)下,避免(mian)形成(cheng)雙(shuang)水(shui)封。光路(lu)與(yu)光強(qiang)度調試(shi)與(yu)檢(jian)查(zha):光強(qiang)度是(shi)測定(ding)靈敏度關鍵(jian)。光強(qiang)度達到(dao)越(yue),靈敏度越(yue)高(gao)。調試(shi)儀(yi)器應選用波(bo)長(chang)於250nm,輻射(she)強度,發光穩(wen)定,對(dui)火焰(yan)狀(zhuang)態反應(ying)遲(chi)鈍的元(yuan)素(su)燈(deng)作為光源(yuan),好(hao)是(shi)銅燈(deng),鎂、鎳(nie)等(deng)元(yuan)素(su)燈(deng)也可(ke)以(yi)。
對(dui)光調(tiao)整(zheng),光源(yuan)對(dui)光調(tiao)試(shi):空(kong)心(xin)陰(yin)極(ji)燈(deng)要在(zai)不(bu)點(dian)亮(liang)情況(kuang)下進行(xing)安裝(zhuang);燈(deng)的前後安裝(zhuang)位置定位(wei)參(can)考如(ru)下:832.1nm空(kong)心(xin)陰(yin)極(ji)燈(deng),其石(shi)英(ying)窗口(kou)距(ju)透(tou)鏡(jing)筒(tong)約(yue)2cm,193.7nm空(kong)心(xin)陰(yin)極(ji)燈(deng),其石(shi)英(ying)窗口(kou)碰(peng)到透(tou)鏡(jing)筒(tong),其(qi)余(yu)元(yuan)素(su)燈(deng)的安裝(zhuang)即按(an)波(bo)長(chang)確(que)定其(qi)前後位(wei)置,波長(chang)值愈愈靠(kao)前。接通(tong)電源,點(dian)燃(ran)元(yuan)素(su)燈(deng)。移動(dong)燈(deng)的位(wei)置(zhi)分別調節(jie)燈(deng)的前後、升(sheng)降(jiang)、旋(xuan)轉(左(zuo)右(you))旋(xuan)鈕,調(tiao)單色(se)器波長(chang)至(zhi)該(gai)元(yuan)素(su)靈敏線(xian),使(shi)顯示(shi)有信(xin)號(hao)輸(shu)出(chu),使(shi)接收器得(de)到(dao)光強(qiang)。檢(jian)查(zha)方法用壹(yi)張白紙(zhi)擋光檢(jian)查(zha),陰極(ji)光斑應聚(ju)焦成像(為正(zheng)圓而不(bu)是(shi)橢(tuo)圓(yuan))其成像位(wei)置在(zai)燃(ran)燒(shao)器縫隙中(zhong)央(yang)或(huo)稍微靠(kao)近單色(se)器壹(yi)方。
燃(ran)燒(shao)器對光調(tiao)試(shi):將(jiang)燃(ran)燒(shao)器縫隙位於光軸之(zhi)下並平(ping)行於光軸,通(tong)過改變燃(ran)燒(shao)器前後、轉(zhuan)角、水平(ping)位置(zhi)來(lai)實現。先(xian)調(tiao)節能量(liang)顯(xian)示(shi)為值,再將用(yong)牙(ya)簽(qian)或火柴(chai)桿插(cha)在(zai)燃(ran)燒(shao)器縫隙中(zhong)央(yang),能(neng)量(liang)顯(xian)示(shi)應(ying)從回(hui)到零。然(ran)後將(jiang)牙簽或火柴(chai)桿垂直(zhi)放置在(zai)縫(feng)隙兩端,顯(xian)示(shi)的透(tou)光度應降(jiang)至(zhi)30%左(zuo)右(you),如(ru)達不(bu)到上述(shu)指標,應(ying)對燃(ran)燒(shao)器的位(wei)置(zhi)再稍微調(tiao)節(jie),直(zhi)到(dao)滿足要(yao)求為止。同時(shi)可(ke)以(yi)點(dian)燃(ran)火焰(yan),噴霧該(gai)元(yuan)素(su)的標(biao)準溶液(ye),調節(jie)燃(ran)燒(shao)器的位(wei)置(zhi),到出(chu)現吸(xi)光度為止。
噴霧器調整,噴霧器中(zhong)的毛(mao)細(xi)管和(he)節(jie)流(liu)嘴(zui)的相(xiang)對(dui)位置和(he)同心度是(shi)調節(jie)噴霧器的關(guan)鍵(jian),毛(mao)細(xi)管口(kou)和(he)節(jie)流(liu)嘴(zui)同心度愈高(gao)、霧滴(di)愈細(xi),霧化(hua)效(xiao)率愈高(gao)。壹(yi)般可(ke)以(yi)通(tong)過觀察噴霧狀(zhuang)況(kuang)來(lai)判斷(duan)調整的效(xiao)果(guo),拆開(kai)噴霧器,用壹(yi)張濾(lv)紙(zhi),將(jiang)霧噴到(dao)濾(lv)紙(zhi)上(shang),濾(lv)紙(zhi)稍濕且非(fei)常(chang)均勻則是(shi)恰(qia)到好處的位(wei)置(zhi)。有些(xie)儀器的噴霧器是(shi)可(ke)調(tiao)的,在(zai)未(wei)點(dian)火時(shi),先(xian)將(jiang)噴霧器調節到反噴位(wei)置(zhi),即插(cha)入(ru)液(ye)面(mian)的毛(mao)細(xi)管出(chu)現氣(qi)泡(pao),然(ran)後點(dian)燃(ran)火焰(yan)噴霧標(biao)準液(ye),按(an)相(xiang)反方向(xiang)慢慢移動(dong),得(de)到(dao)吸(xi)光度便(bian)可(ke)固定(ding)下來(lai)。